Wildeye Wiring: Seametrics FT440
Tools required:
- Hex Key #10
- Wire strippers
- Wrench
Installation Materials:
- 1 x Wildeye Cable (part# WILD-CAB-M12-M-5)
- 1 x Seametrics Cable Gland/Stain Relief (part# 103028-01)
- This should already be attached
WARNING: Seametrics may void the warranty if substitute part/solution is used.
Note: The Seametrics requires an external power source, the wildeye will not work if the meter is not powered
Instructions:
- Unscrew the 4 hex screws on the top lid with #10 Hex key
- Unscrew the Seametrics glad off of the side of the top portion and slide the WILD-CAB-M12-M-5 through the glad. See Picture 1
- Locate the Pulse output connectors on the bottom portion of the meter. See Picture 2.
- Referring to Picture 2, secure the white wire into the “+” side and the black wire into the “-” side
- Give a very gentle tug to ensure the wires are secure
- Secure the top portion of the meter back onto the bottom
- Screw in the Hex Screws and secure with Hex Key
- Tighten the Seametrics gland. Do not over tighten
- Check the waterproof tubing to ensure that the meter is water tight
- Connect the field connector that is on the opposite side of the WILD-CAB-M12-M-5 to the wildeye with the wire that is labeled “C”, keeping in mind there may be other letters labeled on the wire as well.
Picture 1: WILD-CAB-M12-M-5
Picture 2: Bottom portion of the FT440 showing the location of the Pulse Out 1
Pulse Configuration
Note: If the wildeye is attached, using a tackle box method. The Flow meter must be a FTXXX-XX-50
FT4x0 units have a setting called "Set P" which allows one to view or change the volume of flow totalized per pulse sent to "pulse out 1".
"Pulse Out 1" has a maximum scaled pulse output of 6.5Hz, and a fixed pulse width of 100ms.
Therefore, adhere to the following pulse weight selection when configuring the "Set P":
1 Gallon per Pulse |
works for flow rates not more than 360 GPM |
10 Gallon per Pulse |
works for flow rates not more than 3600 GPM |
Pulse out 2 wiring/configuration
If the pulse out 1 is being used, the wildeye should be wired to the Pulse out 2(as shown here, not currently wired in)
- Get into the main menu. Keep everything the same
- Use the > arrow button to move across the menu the the “Exit” tab
- Use the ^ arrow button and press it 4 times
- Scroll to where it says “Out P”, click the ✔️
- Scroll to where it says “Set P2”
- Set this to:
-
0-10 inches Pipe diameter
10 gallons per pulse
10-16 inches Pipe diameter
100 gallons per pulse
- Scroll to the “Exit” tab, click the ✔️
-
Take a picture of the serial number and send an email to supportusa@mywildeye.com. Send:
- Picture of serial number
- Picture of totaliser
- Pulse number used
- Op number on the side of the data logger